![Rohde & Schwarz R&S FSP Series Operating Manual Download Page 577](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864577.webp)
Operating Manual 1164.4556.12 - 06
6.183
R&S FSP
Remote Control – Description of Commands
SENSe Subsystem
This command is only available for model R&S FSP 40 with option R&S FSP-B21 and external mixer
activated.
[SENSe<1|2>:]MIXer:SIGNal
ON | OFF | AUTO | ALL
In analyzer mode, this command activates the Signal ID (ON) or Auto ID (AUTO) mode or both (ALL).
Example:
"MIX ON"
"MIX:SIGN ON"
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
A
This command is only available for model R&S FSP 40 with option R&S FSP-B21 and external mixer
activated.
[SENSe<1|2>:]MIXer[:STATe]
ON | OFF
This command activates or shuts off the external mixer.
Example:
"MIX ON"
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
A
This command is only available for model R&S FSP 40 with option R&S FSP-B21 and external mixer
activated.
[SENSe<1|2>:]MIXer:THReshold
<numeric_value>
This command determines the maximum permissible level difference between measurement and
reference sweep in AUTO ID mode.
Parameter:
<numeric_value> := 0.1 to 100 dB
Example:
"MIX ON"
"MIX:THR 20DB"
Characteristics:
*RST value: 10 dB
SCPI: device-specific
Mode:
A
This command is only available for model R&S FSP 40 with option R&S FSP-B21 and external mixer
activated.
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...