![Rohde & Schwarz R&S FSP Series Operating Manual Download Page 601](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864601.webp)
Operating Manual 1164.4556.12 - 06
6.207
R&S FSP
Remote Control – Description of Commands
SOURce Subsystem
6.17 SOURce Subsystem
The SOURce subsystem controls the output signals of the instrument if the options tracking generator
(B9) or External Generator Control (B10) are installed. The measurement window is selected by
SOURce1 (screen A) and SOURce2 (screen B).
6.17.1
Internal Tracking Generator
SOURce<1|2>:AM:STATe
ON | OFF
This command switches on or off the external amplitude modulation of the tracking generator in the
selected measurement window.
External I/Q modulation is switched off, if active. This command is only valid in conjunction with the
tracking generator option B9.
Example:
"SOUR:AM:STAT ON"
'Switches on the external amplitude modulation of the tracking generator for
screen A.
Characteristics:
*RST value: OFF
SCPI: conform
Mode:
A
SOURce<1|2>:DM:STATe
ON | OFF
This command switches on or off the external I/Q modulation of the tracking generator in the selected
measurement window.
External AM and external FM are switched off, if active. This command is only valid in conjunction with
the tracking generator option B9.
Example:
"SOUR2:DM:STAT ON"
'Switches on the external I/Q modulation of the tracking generator for screen B.
Characteristics:
*RST value: OFF
SCPI: conform
Mode:
A
SOURce<1|2>:FM:DEViation
100Hz to 10MHz
This command defines the maximum frequency deviation at 1 V input voltage at the FM input of the
tracking generator.
The possible value range is 100 Hz to 10 MHz in steps of one decade.
This command is only available in connection with the Tracking Generator Option B9.
Example:
"SOUR:FM:DEV 1MHz"
'Sets the maximum frequency deviation of the tracking generator for screen A to 1
MHz.
Characteristics:
*RST value: 100 Hz
SCPI: conform
Mode:
A
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...