![Rohde & Schwarz R&S FSP Series Operating Manual Download Page 254](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864254.webp)
R&S FSP
Instrument Functions
Basic Settings
4.182
Operating Manual 1164.4556.12 - 06
With the 'Auto Login' option active, an automatic registration is performed during
booting with the specified user name and password. Otherwise the Windows XP
login request is displayed during booting.
Remote command:
--
Enabling Firmware Options
The
OPTIONS
softkey opens a submenu that allows license keys for firmware
options to be entered. Previously installed options are displayed in a table that
opens automatically.
Due to the additional available applications the required resources may exceed the
available system memory on instruments shipped with a system memory size below
1 GByte. Most applications require additional memory space once activated. These
resources are not released when the application is quit to allow a fast re-activation
of this application for measurement performance reasons.
As of firmware version 4.5x it is possible to manually enable / disable a license key
code once entered, if the required system memory would exceed the available sys-
tem memory. The application hotkey will not be visible in the hotkey bar if the license
key is disabled.
The MIN MEM value lists the additional base system memory required for this appli-
cation. This part of the memory remains allocated when the application is left and
therefore is not available for other applications. The MAX MEM value is the addi-
tional worst case memory demand of the application.
The required system memory is the sum of the enabled applications MIN MEM val-
ues plus the maximum value of all enabled application's MAX MEM values.
Some applications can not be disabled. The cursor will skip these entries. A few
applications are combined to groups and they can only be enabled or disabled as a
whole.
If it's not possible to enable a required application there are to ways to save system
memory:
A PC keyboard with trackball (or mouse instead) is required for the installation/
configuration of the network support.
The softkey is only available with built-in LAN interface R&S FSP-B16.
For information on deactivating/activating the auto login mechanism see section
“Logging on to the Network“ in chapter ’LAN Interface’ of the Quick Start Guide.
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...