![Rohde & Schwarz R&S FSP Series Operating Manual Download Page 177](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864177.webp)
Operating Manual 1164.4556.12 - 06
4.105
R&S FSP
Instrument Functions
Analyzer Mode
Some of the measurement specifications (e.g. PDC, RCR STD-27B) require mea-
surement of the occupied bandwidth using a peak detector. The detector setting of
the R&S FSP has to be changed accordingly then.
Example
Measurement of occupied bandwidth of a PDC signal at 800 MHz, level 0 dBm
[
PRESET
]
Set the R&S FSP to the default setting.
[
FREQ
:
CENTER
: 800
MHz]
Set the center frequency to 800 MHz.
[
AMPT
: 0 dBm]
Set the reference level to 0 dBm.
[
MEAS
]
Call the menu for the measurement functions.
[
OCCUPIED
BANDWIDTH
]
Select measurement of the occupied bandwidth and open
the submenu for configuring the measurement.
[% POWER
BANDWIDTH
: 99
%
]
Select 99% for the bandwidth to be measured.
[
CHANNEL
BANDWIDTH
: 21
kHz
]
Enter the channel bandwidth of 21 kHz specified by PDC.
[
ADJUST SETTINGS
]
Optimize the measurement parameters for the specified
channel bandwidth.
Allow for a complete frequency sweep so that the
R&S FSP can determine the total signal power.
[
ADJUST REF LVL
]
Adjust the reference level to the measured signal power.
[
TRACE
:
DETECTOR
:
PDC requires measurement of the occupied bandwidth
using a
DETECTOR MAX PEAK
] peak detector. Therefore, switch on the peak detector
instead of the RMS detector selected by
ADJUST SET-
TINGS
.
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...