Filter edge of a 300 kHz filter used as an FM-discriminator characteristic
5. Set an FM deviation of 100 kHz and an AF of 1 kHz on the signal generator
6. Set a frequency deviation of 0 Hz on the R&S FSP
➢
Press the SPAN key.
➢
Press the ZERO SPAN.
The demodulated FM signal is displayed. The signal moves across the screen.
7. Creating a stable display by video triggering
➢
Press the TRIG key.
➢
Press the VIDEO softkey.
A stationary display is obtained for the FM AF signal
Result: (-10 ± 5) dB; this means that a deviation of 100 kHz is obtained if the demod-
ulator characteristic slope is 5 dB/100 kHz
Fig. 2.18 Demodulated FM signal