Operating Manual 1164.4556.12 - 06
NEXT PEAK
The
NEXT PEAK
softkey sets the active marker/delta marker to the next lower max-
imum of the selected trace.
Remote command:
CALC:MARK:MAX:NEXT
CALC:DELT:MAX:NEXT
NEXT PEAK
RIGHT
The
NEXT PEAK RIGHT
softkey sets the active marker/delta marker to the next
lower maximum right of the current marker position on the selected trace.
Remote command:
CALC:MARK:MAX:RIGH
CALC:DELT:MAX:RIGH
NEXT PEAK
LEFT
The
NEXT PEAK LEFT
softkey sets the active marker/delta marker to the next lower
maximum left of the current marker position the selected trace.
Remote command:
CALC:MARK:MAX:LEFT
CALC:DELT:MAX:LEFT
SEARCH
LIMITS
The
SEARCH LIMITS
softkey limits the search range for maximum or minimum
search. The softkey switches to a submenu in which the search range limits can be
set in the x and y direction.
LEFT LIMIT /
RIGHT LIMIT
The
LEFT LIMIT
and
RIGHT LIMIT
softkeys define the two vertical lines F1 and F2
in the frequency domain (span > 0) and T1 / T2 in the time domain (span = 0). The
search is performed between these lines in the frequency and time domain
If only
LEFT LIMIT
is enabled, line F1/T1 is the lower limit and the upper limit corre-
sponds to the stop frequency. If
RIGHT LIMIT
is also enabled, it determines the
upper limit.
Remote command:
CALC:MARK:X:SLIM:LEFT 1MHZ
CALC:MARK:X:SLIM:RIGH 10MHZ
CALC:MARK:X:SLIM ON
THRESHOLD
The
THRESHOLD
softkey defines the threshold line.
The threshold line represents a limit for the level range of the max. search at the
lower end and that of the min. search at the upper end.
Remote command:
CALC:THR -20dBm
CALC:THR ON
[
REF LEVEL = MKR LVL
] The reference level is set to the measured marker level.
[
SPAN
]
The span can, for example, be reduced using the rotary
knob.
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...