CALCulate<1|2>:LIMit<1...8>:LOWer:OFFSet
<numeric_value>
This command defines an offset for the y-axis of the selected relative lower limit line. In contrast to
CALC:LIM:LOW:SHIFt
, the line is not shifted by modifying the individual Y values but by means of
an additive offset. The offset is independent of the measurement window.
Example:
"CALC:LIM2:LOW:OFFS 3dB"
'Shifts limit line 2 in the corresponding measurement windows by 3 dB upwards.
Characteristics:
*RST value: 0
SCPI: device-specific
Mode:
A
CALCulate<1|2>:LIMit<1...8>:LOWer:SHIFt
<numeric_value>
This command shifts a limit line by the indicated value in Y direction. In contrast to
CALC:LIM:LOW:
OFFS
, the line is shifted by modifying the individual Y values but not by means of an additive offset.
Up to 8 limit lines can be defined at the same time. The shift is independent of the measurement
window.
Example:
"CALC:LIM3:LOW:SHIF 20DB"
'Shifts all Y values of limit line 3 by 20 dB.
Characteristics:
*RST value: --
SCPI: device-specific
Mode:
A
This command is an event and therefore has no *RST value and no query.
CALCulate<1|2>:LIMit<1...8>:LOWer:SPACing
LINear | LOGarithmic
This command selects linear or logarithmic interpolation for the lower limit line.
Example:
"CALC:LIM:LOW:SPAC LIN"
Characteristics:
*RST value: LIN
SCPI: device-specific
Mode:
A
CALCulate<1|2>:LIMit<1...8>:LOWer:STATe
ON | OFF
This command switches on or off the indicated limit line. The limit check is activated separately with
CALC:LIM:STAT ON.
Up to 8 limit lines can be defined at the same time. The numeric suffixes <1|2> of CALCulate indicate
the measurement window.
In analyzer mode, the result of the limit check can be queried with
CALCulate:LIMit<1...8>:
FAIL?
.
Example:
"CALC:LIM4:LOW:STAT ON"
'Switches on limit line 4 (lower limit) in screen A.
"CALC2:LIM4:LOW:STAT ON"
'Switches on limit line 4 (lower limit) also in screen B.
Characteristics:
*RST value: OFF
SCPI: conform
Mode:
A
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...