Operating Manual 1164.4556.12 - 06
4.69
R&S FSP
Instrument Functions
Analyzer Mode
SELECT
MARKER
The
SELECT MARKER
softkey activates the numerical selection of the marker in
the data entry field. Delta marker 1 is selected by input of ' 0 '.
Remote command:
CALC:MARK1 ON
CALC:MARK1:X <value>
CALC:MARK1:Y?
PEAK
The
PEAK
softkey sets the active marker/delta marker to the peak of the trace. If no
marker is active when
MKR->
menu is called, MARKER 1 is automatically switched
on and the peak search is performed.
Remote command:
CALC:MARK:MAX
CALC:DELT:MAX
CENTER =
MKR FREQ
The
CENTER = MKR FREQ
softkey sets the center frequency to the current marker
or delta marker frequency.
A signal can thus be set to the center of the frequency display range, for example,
so that it can then be examined in detail with a smaller span.
The softkey is not available in the time domain (zero span).
Remote command:
CALC:MARK:FUNC:CENT
Example
A spectrum is displayed with a large span after PRESET. A signal off the center is to
be examined in detail:
REF LEVEL =
MKR LVL
The
REF LEVEL = MKR LVL
softkey sets the reference level to the current marker
level.
Remote command:
CALC:MARK:FUNC:REF
Example
A spectrum is displayed with a large span after PRESET. A signal off the center is to
be examined in detail:
[
PRESET
]
R&S FSP is set to the default setting.
[
MKR->
]
MARKER 1 is switched on and automatically jumps to
the largest signal of the trace.
[
CENTER=MKR FREQ
]
The center frequency is set to the marker frequency. The
span is adapted in such a way that the minimum fre-
quency (= 0 Hz) or the maximum frequency is not
exceeded.
[
SPAN
]
The span can, for example, be reduced using the rotary
knob.
[
PRESET
]
R&S FSP is set to the default setting.
[
MKR->
]
MARKER 1 is switched on and automatically jumps to
the largest signal of the trace.
[
CENTER=MKR FREQ
]
The center frequency is set to the marker frequency.
The span is adapted in such a way that the minimum fre-
quency (= 0 Hz) or the maximum frequency is not
exceeded.
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...