![Rohde & Schwarz R&S FSP Series Operating Manual Download Page 496](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864496.webp)
R&S FSP
Remote Control – Description of Commands
DIAGnostic Subsystem
6.102
Operating Manual 1164.4556.12 - 06
6.7 DIAGnostic Subsystem
The DIAGnostic subsystem contains the commands which support instrument diagnostics for mainte-
nance, service and repair. In accordance with the SCPI standard, all of these commands are device-spe-
cific.
The measurement windows are selected by DIAGnostic1 (screen A) or DIAGnostic2 (screen B).
DIAGnostic<1|2>:SERVice:CSOource[:POWer]
<numeric_value>
This command switches the level of the 128 MHz reference signal source between 0 dBm and -30 dBm
in the selected measurement window.
Example:
"DIAG:SERV:CSO 0DBM"
Characteristics:
*RST value: -30 dBm
SCPI: device-specific
Mode:
A
DIAGnostic<1|2>:SERVice:HWINfo?
This command queries the contents of the module info table. Table lines are output as string data and
are separated by commas.
"<component 1>|<serial #>|<order #>|<model>|<HWC>|<rev>|<sub rev>",
"<component 2>|<serial #>|<order #>|<model>|<HWC>|<rev>|<sub rev>", ...
The individual columns of the table are separated from each other by '|'.
The numeric suffix <1|2> is ignored with this command.
Example:
"DIAG:SERV:HWIN?"
Result (shortened):
"RF_ATTEN_7|650551/007|1067.7684|02|00|20|04",
"IF-FILTER|648158/037|1093.5540|03|01|07|05",
...
Characteristics:
*RST value: --
SCPI: device-specific
Mode:
A
DIAGnostic<1|2>:SERVice:INPut:PULSed:PRATe
10 kHz | 62.5 kHz | 640 MHz
This command selects the pulse rate for the pulsed calibration signal in the selected measurement
window.
Example:
"DIAG:SERV:INP:PULS:PRAT 62.5 kHz"
Characteristics:
*RST value: 62.5 kHz (with option R&S FSP-B15), 640 MHz (R&S FSP40 without
option R&S FSP-B15)
SCPI: device-specific
Mode:
A
This command is only available in conjunction with the optional Broadband Calibration Source
R&S FSP-B15.
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...