R&S FSP
Remote Control – Description of Commands
SENSe Subsystem
6.174
Operating Manual 1164.4556.12 - 06
[SENSe<1|2>:]LIST:POWer:SET
<PEAK meas>,<RMS meas>,<AVG meas>,<trigger mode>,<trigger
slope>,<trigger offset>, <gate length>
This command defines the constant settings for the list during multiple power measurement.
Parameters <PEAK meas>, <RMS meas> and <AVG meas> define, which measurements are to be
performed at the same time at the frequency point. Correspondingly, one, two or three results per
frequency point are returned for the
SENS:LIST:POW?
command. If all three parameters are set to
OFF, the command generates an execution error.
Parameter:
<PEAK meas>: ON activates the measurement of the peak power (peak detector).
OFF deactivates the measurement of the peak power.
<RMS meas>: ON activates the measurement of the RMS power (RMS detector).
OFF deactivates the measurement of the RMS power.
<AVG meas>: ON activates the measurement of the average power (average
detector). OFF deactivates the measurement of the average power.
<trigger mode>: Selection of the trigger source used for the list measurement.
Possible values: IMMediate | VIDeo | EXTernal | IFPower
<trigger slope>: Used trigger slope.
Possible values: POSitive | NEGative
<trigger offset>: Offset between the detection of the trigger signal and the start of
the measurement at the next frequency point.
Range of values: 0 s, 125 ns to 100s
<gate length>: Gate length with Gated Sweep.
Range of values: 0 s, 125 ns to 100s
Return value:
The query command returns a list of comma-separated values (CSV) of the
settings, i.e.
ON,ON,ON,IMM,POS,0,0
if the configuration has been set with the command
"SENSe:LIST:POWer:SET ON,ON,ON,IMM,POS,0,0"
Example:
"SENSe:LIST:POWer:SET ON,OFF,OFF,EXT,POS,10US,434US"
"SENSe:LIST:POWer:SET ON,ON,ON,VID,NEG,10US,0"
Characteristics:
*RST values: ON,OFF,OFF,IMM,POS,0S,0S
SCPI: device-specific
Mode:
A
• The value 0 s deactivates the use of GATED TRIGGER; other values activate
the GATED TRIGGER function.
• Values <> 0 s are only possible if <trigger mode> is different from IMMediate.
Otherwise, an execution error is triggered.
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...