R&S FSP
Remote Control – Description of Commands
SENSe Subsystem
6.16.12
SENSe:SWEep Subsystem
This subsystem controls the sweep parameters.
The measurement windows are selected by SENSe1 (screen A) and SENSe2 (screen B).
[SENSe<1|2>:]SWEep:COUNt
0 to 32767
This command defines the number of sweeps started with single sweep, which are used for calculating
the average or maximum value. In average mode, the value 0 defines a continuous averaging of
measurement data over 10 sweeps.
Example:
"SWE:COUN 64"
'Sets the number of sweeps to 64.
"INIT:CONT OFF"
'Switches to single-sweep mode.
"INIT;*WAI"
'Starts a sweep and waits for its end.
Characteristics:
*RST value: 0
SCPI: conform
Mode:
A
[SENSe<1|2>:]SWEep:COUNt:CURRent?
This query command returns the current number of started sweeps. A sweep count value should be
set and the device should be in single-sweep mode. This command is a query and therefore has no
*RST value.
Example:
"SWE:COUNt 64"
'sets sweep count to 64
"INIT:CONT OFF"
'switches to single-sweep mode
"INIT"
'starts a sweep (without waiting for the sweep end!)
"SWE:COUN:CURR?"
'queries the number of started sweeps
Characteristics:
*RST value: --
SCPI: conform
Mode:
A
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...