![Rohde & Schwarz R&S FSP Series Operating Manual Download Page 484](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864484.webp)
R&S FSP
Remote Control – Description of Commands
CALCulate Subsystem
6.90
Operating Manual 1164.4556.12 - 06
6.5.7
CALCulate:MATH Subsystem
The CALCulate:MATH subsystem allows to process data from the SENSe-subsystem in numeric expres-
sions.
The measurement windows are selected by CALCulate1 (screen A) and CALCulate2 (screen B).
CALCulate<1|2>:MATH[:EXPression][:DEFine]
(<expr>)
This command defines the mathematical expression for relating traces to trace1.
The
command switches the mathematical relation of traces on or off.
The measurement windows are selected via CALCulate1 (screen A) or 2 (screen B).
Parameter:
<expr>::= 'OP1 - OP2'
OP1 ::= TRACE1
OP2 ::= TRACE2 | TRACE3
Example:
"CALC1:MATH (TRACE1 - TRACE2)"
'Selects the subtraction of trace 1 from trace 2 in screen A.
"CALC2:MATH (TRACE1 - TRACE3)"
'Selects the subtraction of trace 1 from trace 3 in screen B.
Characteristics:
*RST value: -
SCPI: conform
Mode:
A-T
CALCulate<1|2>:MATH:MODE
LINear | LOGarithmic | POWer
This command selects linear or logarithmic (= video) calculation of the mathematical functions related
to the traces. The calculation of the average is one of the affected functions. The setting is valid for all
measurement windows, i.e. the numeric suffix
<1|2>
of
CALCulate
is irrelevant.
Example:
"CALC:MATH:MODE LIN"
'Switches on the linear calculation.
Characteristics:
*RST value: LOG
SCPI: device-specific
Mode:
A-T
CALCulate<1|2>:MATH:POSition
-100PCT to 200PCT
This command defines the position of the result of the trace mathematics in the selected measurement
window. The indication is in % of the screen height, with 100% corresponding to the upper diagram
border.
Example:
"CALC:MATH:POS 50PCT"
'Sets the position in screen A to the horizontal diagram center.
Characteristics:
*RST value: 50 %
SCPI: device-specific
Mode:
A-T
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...