![Rohde & Schwarz R&S FSP Series Operating Manual Download Page 580](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864580.webp)
R&S FSP
Remote Control – Description of Commands
SENSe Subsystem
6.186
Operating Manual 1164.4556.12 - 06
[SENSe<1|2>:]MPOWer:RESult:MIN?
This command queries the minimum power value in a multiple burst power measurement as
configured and initiated with
SENSe:MPOWer[:SEQuence]
. The unit used for the return values is
always dBm.
If no measurement result is available, the command will return a query error.
Example:
*ESE 1
*SRE 32
'Configuration of status reporting systems for the generation of an SRQ on
operation complete
SENSe:MPOWer
935.2MHz,1MHz,434us,VIDEO,50PCT,5us,MEAN,20;*OPC
'Configuring and starting the measurement
...
'Further actions of the control computer during measurement
On SRQ:
'Response to service request
SENSe:MPOWer:RESult:MIN?
Characteristics:
*RST value: --
SCPI: instrument-specific
Mode:
A-F, A-T
[SENSe<1|2>:]MPOWer[:SEQuence]
<analyzer freq>,<rbw>,<meas time>,<trigger source>,
<trigger level>,<trigger offset>,<type of meas>,<# of meas>
This command configures the instrument setup for multiple burst power measurement and starts a
measurement sequence. When synchronizing the command with *OPC, a service request is
generated as soon as the defined number of individual measurements (# of meas) is reached.
To reduce the setting time, the setup is performed simultaneously for all selected parameters.
The command in the form of a query makes the instrument settings, performs the defined number of
measurements and outputs the measurement results list.
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...