R&S FSP
Instrument Functions
Analyzer Mode
4.40
Operating Manual 1164.4556.12 - 06
where
The display "Sweep N of N" does not change any more until a new start is triggered.
In the
SINGLE SWEEP
mode, the number of sweeps is triggered with
SWEEP
START
. The sweeps are stopped when the selected number of sweeps is attained.
The number of the current sweep and the total number of sweeps are shown on the
display: "Sweep 3 of 200".
VIEW
The
VIEW
softkey freezes the current contents of the trace memory and displays it.
If a trace is frozen by
VIEW
, the instrument settings can be changed without the dis-
played trace being modified (exception: level display range and reference level, see
below). The fact that the trace and the current instrument setting do not agree any
more is indicated by an enhancement label "*" at the right edge of the grid.
If in the
VIEW
display mode the level display range (
RANGE
) or the reference level
(
REF LEVEL
) are changed, R&S FSP automatically adapts the measured data to
the changed display range. This allows an amplitude zoom to be made after the
measurement in order to show details of the trace.
Remote command:
DISP:WIND:TRAC:MODE VIEW
BLANK
The
BLANK
softkey activates the blanking of the trace on the screen.
Remote command:
DISP:WIND:TRAC OFF
SWEEP
COUNT
The
SWEEP COUNT
softkey activates the entry of the number of sweeps used for
averaging. The allowed range of values is 0 to 30000 and the following should be
observed:
• Sweep Count = 0 means running averaging
• Sweep Count = 1 means no averaging, maxhold or minhold is carried out
• Sweep Count > 1 means averaging over the selected number of sweeps; in the
continuous sweep mode averaging is performed until the set number of sweeps
is attained and is then continued as running averaging.
The default setting is running averaging (Sweep Count = 0). The number of sweeps
used for averaging is the same for all active traces in the selected diagram.
Remote command:
SWE:COUN 64
DETECTOR
See following section
“Selection of Detector” on page 4.44
.
Trace =
new
trace
Trace
old
= old trace
N
= SWEEP COUNT
Trace
N 1
–
(
)
Trace
old
⋅
MeasValue
+
N
---------------------------------------------------------------------------------
=
The setting of the sweep count in the
TRACE
menu is equivalent to the setting in
the
SWEEP
menu.
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...