4.9.2
Transmission Measurement
The transmission characteristic of a two-port network is measured. The external
generator serves as a signal source. It is connected to the input connector of the
DUT. The input of the R&S FSP is fed from the output of the DUT.
Fig. 4.28 Test setup for transmission measurement
A calibration can be carried out to compensate for the effects of the test setup (e.g.
frequency response of connecting cables).
4.9.2.1
Calibration of Transmission Measurement
SOURCE CAL
The
SOURCE CAL
softkey opens a submenu comprising the calibration functions
for the transmission and reflection measurement.
For information on the calibration of the reflection measurement
(
CAL REFL SHORT
and
CAL REFL OPEN
) and its mechanisms, refer to section
“Calibration of Reflection Measurement” on page 4.234
.
To carry out a calibration for transmission measurements the whole test setup is
through-connected (THRU).
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...