![Rohde & Schwarz R&S FSP Series Operating Manual Download Page 148](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864148.webp)
R&S FSP
Instrument Functions
Analyzer Mode
4.76
Operating Manual 1164.4556.12 - 06
• Spurious emissions (
“SPURIOUS EMISSIONS” on page 4.122
)
• Spectrum Emission Mask (
“SPECTRUM EMISSION MASK” on page 4.140
• 3
rd
order intercept (
)
• Harmonic distortion (
“HARMONIC DISTOR” on page 4.120
The above measurements are carried out alternatively.
4.5.13.1
Power Measurement in Time Domain
With the aid of the power measurement function, the R&S FSP determines the
power of the signal in the time domain (SPAN = 0 Hz) by summing up the power at
the individual pixels and dividing the result by the number of pixels. In this way it is
possible to measure for example the power of TDMA signals during transmission or
during the muting phase. Both the mean power and the rms power can be measured
by means of the individual power values.
The result is displayed in the marker info field.
The measured values are updated after each sweep or averaged over a user-
defined number of sweeps (
AVERAGE ON/OFF
and
NUMBER OF SWEEPS
) in
order to determine e.g. the mean power over several bursts. For determination of
the peak value (
MAX HOLD ON
) the maximum value from several sweeps is dis-
played.
Example
Marker info field for:
MEAN
selected,
AVERAGE ON
and
MAX HOLD ON
:
If both the on and off phase of a burst signal are displayed, the measurement range
can be limited to the transmission or to the muting phase with the aid of vertical
lines. The ratio between signal and noise power of a TDMA signal for instance can
be measured by using a measurement as a reference value and after that varying
the measurement range.
Upon switching on power measurement the sample detector is activated (
TRACE -
DETECTOR - SAMPLE)
.
TIME DOM
POWER
MEAN HOLD -2.33 dBm
MEAN AV -2.39 dBm
Side menu
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...