Operating Manual 1164.4556.12 - 06
4.231
R&S FSP
Instrument Functions
External Generator Control – Option R&S FSP-B10
Fig. 4.31 Normalized measurement, shifted with REF VALUE POSITION 50%
Remote command:
DISP:WIND:TRAC:Y:RPOS 10PCT
REF VALUE
The
REF VALUE
softkey activates the input of a level difference which is assigned
to the reference line.
In the default setting, the reference line corresponds to a level difference of 0 dB. If
e.g. a 10-dB attenuator pad is inserted between DUT and R&S FSP input between
recording of the calibration data and normalization, the trace will be shifted down by
10 dB. By entering a
REF VALUE
of –10 dB the reference line for difference forma-
tion can also be shifted down by 10 dB so that it will again coincide with the trace
(see
).