Operating Manual 1164.4556.12 - 06
4.239
R&S FSP
Instrument Functions
External Generator Control – Option R&S FSP-B10
• TYPE
The field opens the list with the available generators:
After completion of the selection, the remaining fields of the table are filled with
the generator characteristics.
A list of generator types supported by the R&S FSP is to be found in section
of Generator Types Supported by the R&S FSP” on page 4.240
• IFC
This field selects the interface type of external generator 1 or 2. The following
types are available:
–
GPIB
: GPIB only, suitable for all generators of other manufacturers and some
Rohde & Schwarz instruments
–
TTL
: GPIB and TTL interface for synchronization, for most of the Rohde &
Schwarz generators, see table above.
The two operating modes differ in the speed of the control: Whereas, with pure
GPIB operation, each frequency to be set must be individually transferred to the
generator, additional use of the TTL interface permits to program a total frequency
list at once and subsequently perform the frequency stepping via TTL handshake,
which is a big advantage in terms of speed.
•
GPIB
ADDR
GPIB address of the respective generator. Addresses from 0 to 30 are possible.
• MODE
Operating mode of generator. The generator activated using the
FREQUENCY
SWEEP
softkey is automatically set to remote mode (REMOTE), the other to
manual operation (LOCAL).
•
F MIN
F MAX
Frequency range of generator. Select the start and stop frequency of the
R&S FSP in a way that the specified range is not exceeded. If the start frequency
lies below
F MIN
, the generator is only switched on when
F MIN
is reached. If the
stop frequency lies above F MAX, it is limited to F MAX when the generator is
switched on using the
EXT SRC ON/OFF
softkey.
•
P MIN
P MAX
Level range of generator. This field defines the allowed input range for the
POWER
column in the
FREQUENCY SWEEP
table.
Remote command:
SYST:COMM:RDEV:GEN2:TYPE 'SMA01A'
SYST:COMM:RDEV:GEN:LINK TTL
SYST:COMM:GPIB:RDEV:GEN1:ADDR 28
Generators equipped with the TTL interface can also be operated with GPIB only.
Only one of the two generators can be operated with TTL interface at a time. The
other generator must be configured for GPIB.
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...