Tracking Generator – Option R&S FSP-B9
4.212
Operating Manual 1164.4556.12 - 06
4.8 Tracking Generator – Option R&S FSP-B9
During normal operation (without a frequency offset), the tracking generator emits a
signal exactly at the input frequency of the R&S FSP.
For frequency-converting measurements it is possible to set a constant frequency
offset of ±150 MHz between the receive frequency of the R&S FSP and the output
signal of the tracking generator.
Moreover, an I/Q modulation or AM and FM modulation of the output signal can be
provided using two analog input signals.
The output power is level-controlled and can be set in 0.1 dB steps in a range from -
30 dBm to 0 dBm.
The tracking generator can be used in all operating modes. Acquisition of test setup
calibration values (SOURCE CAL) and normalization using these correction values
(
NORMALIZE
) is only possible in the
NETWORK
operating mode.
The tracking generator is activated by means of the
NETWORK
hotkey in the hotkey
bar at the bottom of the screen (for details refer to section
.)
4.8.1
Tracking Generator Settings
The
NETWORK
hotkey opens a menu for selecting the functions of the tracking
generator.
NETWORK
The RF characteristics of some DUTs is especially sensitive concerning the input
VSWR. In such cases insertion of 20 dB attenuation between the DUT and the
tracking generator output is highly recommended.
!
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...