Operating Manual 1164.4556.12 - 06
6.83
R&S FSP
Remote Control – Description of Commands
CALCulate Subsystem
Characteristics:
*RST value: ABSolute
SCPI: device-specific
Mode:
A-T
CALCulate<1|2>:MARKer<1...4>:FUNCtion:SUMMary:PHOLd
ON | OFF
This command switches on or off the peak-hold function for the active time domain power
measurement in the indicated measurement window. The function is independent of the marker
selection, i.e. the numeric suffix <1...4> of
:MARKer
is irrelevant. It is only available in the time domain
(span = 0). The peak-hold function is reset by switching it off and on again.
Example:
"CALC:MARK:FUNC:SUMM:PHOL ON"
'Switches on the function in screen A.
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
A-T
The peak-hold function is reset by switching off and on, again.
CALCulate<1|2>:MARKer<1...4>:FUNCtion:SUMMary:PPEak:AVERage:RESult?
This command is used to query the result of the measurement of the averaged positive peak value in
the selected measurement window. The query is only possible if averaging has been activated
previously using
CALCulate<1|2>:MARKer<1...4>:FUNCtion:SUMMary:AVERage
.
The function is independent of the marker selection, i.e. the numeric suffix <1...4> in MARKer is
irrelevant. It is only available in the time domain (span = 0).
A complete sweep with synchronization to sweep end must be performed between switching on the
function and querying the measured value to obtain a valid query result. This is only possible in single-
sweep mode.
Example:
"INIT:CONT OFF"
'Switches to single-sweep mode.
"CALC:MARK:FUNC:SUMM:PPE ON"
'Switches on the function in screen A.
"CALC:MARK:FUNC:SUMM:AVER ON"
'Switches on the calculation of average in screen A.
"INIT;*WAI"
'Starts a sweep and waits for the end.
"CALC:MARK:FUNC:SUMM:PPE:AVER:RES?"
'Outputs the result of screen A.
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
A-T
This command is only a query and therefore has no *RST value.
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...