![Rohde & Schwarz R&S FSP Series Operating Manual Download Page 557](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864557.webp)
Operating Manual 1164.4556.12 - 06
6.163
R&S FSP
Remote Control – Description of Commands
SENSe Subsystem
[SENSe<1|2>:]ESPectrum:RANGe<1..20>:INSert
AFTer | BEFore
This command inserts a new range before or after the specified range. The range numbers are
updated accordingly. The numeric suffix at RANGe<1...20> specify the range.
The numeric suffixes <1...20> specify the range. The numeric suffix at SENSe<1|2> is irrelevant.
Example:
"ESP:RANG3:INS BEF"
'Inserts a new range before Range 3
"ESP:RANG1:INS AFT"
'Inserts a new range after Range 1
Characteristics:
*RST value:-
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]ESPectrum:RANGe<1..20>:LIMit:ABSolute:STARt
<numeric_value>
This command defines the absolute limit at the start frequency of the range for the Spectrum Emission
Mask measurement.
The numeric suffixes <1...20> specify the range. The numeric suffix at SENSe<1|2> is irrelevant.
Parameter:
-400 to +400 dBm
Example:
"ESP:RANG5:LIM:ABS:STAR -40"
'Absolute limit at start frequency in range 5 is -40 dBm
Characteristics:
*RST value: -13dBm
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]ESPectrum:RANGe<1..20>:LIMit:ABSolute:STOP
<numeric_value>
This command defines the absolute limit at the stop frequency of the range for the Spectrum Emission
Mask measurement.
The numeric suffixes <1...20> specify the range. The numeric suffix at SENSe<1|2> is irrelevant.
Parameter:
-400 to +400 dBm
Example:
"ESP:RANG5:LIM:ABS:STOP -40"
'Absolute limit at stop frequency in range 5 is -40 dBm
Characteristics:
*RST value: -13dBm
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]ESPectrum:RANGe<1..20>:LIMit:RELative:STARt
<numeric_value>
This command defines the relative limit at the start frequency of the range for the Spectrum Emission
Mask measurement.
The numeric suffixes <1...20> specify the range. The numeric suffix at SENSe<1|2> is irrelevant.
Parameter:
-400 to +400 dBc
Example:
"ESP:RANG5:LIM:REL:STAR -40"
'Relative limit at start frequency in range 5 is -40 dBc
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...