![Rohde & Schwarz R&S FSP Series Operating Manual Download Page 116](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864116.webp)
R&S FSP
Instrument Functions
Analyzer Mode
4.44
Operating Manual 1164.4556.12 - 06
Example - data section of the file
DECIM SEP
The
DECIM SEP
softkey selects the decimal separator between '.' (decimal point)
and ',' (comma) with floating-point numerals for the
function.
With the selection of the decimal separator different language versions of evaluation
programs (e.g. Microsoft Excel) can be supported.
Remote command:
FORM:DEXP:DSEP POIN
COPY TRACE
The
COPY TRACE
softkey copies the screen contents of the current trace into
another trace memory. The desired memory is selected by entering the number 1, 2
or 3.
Upon copying, the contents of the selected memory is overwritten and the new con-
tents displayed in view mode.
Remote command:
TRAC:COPY TRACE1,TRACE2
4.5.8.1
Selection of Detector
The detectors of the R&S FSP are implemented as pure digital devices. The following
detectors are available:
•
“Peak detectors (MAX PEAK and MIN PEAK)” on page 4.45
•
“Auto peak detector” on page 4.45
•
“Sample detector” on page 4.45
•
•
“Average detector” on page 4.46
•
“Quasipeak detector” on page 4.46
The peak detectors compare the current level value with the maximum or minimum
level from the previously sampled data. When the number of samples defined by the
instrument setting is reached, the samples are combined in the displayed pixels.
File contents
Description
Trace 1:;;
Selected trace
x-Unit;Hz;
Unit of x values:
Hz with span > 0; s with span = 0;
dBm/dB with statistics measurements
y-Unit;dBm;
Unit of y values:
dB*/V/A/
Ω
depending on the selected unit with y axis LOG or % with
y-axis LIN
Preamplifier;OFF;
Preamplifier state (OFF, if no preamplifier is available)
Transducer;OFF;
Transducer state:
OFF:No transducer active (or available)
<trd name>selected transducer
Values; 625;
Number of measurement points
10000;-10.3;-15.7
10180;-11.5;-16.9
10360;-12.0;-17.4
...;...;
Measured values:
<x value>, <y1>, <y2>
<y2> being available only with detector AUTOPEAK and containing
in this case the smallest of the two measured values for a test point.
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...