![Rohde & Schwarz R&S FSP Series Operating Manual Download Page 779](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864779.webp)
Operating Manual 1164.4556.12 - 06
9.5
R&S FSP
Error Messages
SCPI-Specific Error Messages
-158
String data not allowed
The command contains a valid string at a position which is not allowed.
Example: A text parameter is set in quotation marks,
INPut:COUPling "DC"
-160
Block data error
The command contains faulty block data.
-161
Invalid block data
The command contains faulty block data.
Example: An END message was received prior to reception of the expected number
of data.
-168
Block data not allowed
The command contains valid block data at an invalid position.
Example: The command
*RCL
requires a number to be indicated.
-170
Expression error
The command contains an invalid mathematical expression.
-171
Invalid expression
The command contains an invalid mathematical expression.
Example: The expression contains mismatching parentheses.
-178
Expression data not allowed
The command contains a mathematical expression at an invalid position.
Error code
Error text in the case of queue poll
Error explanation
-200
Execution error
Error on execution of the command.
-201
Invalid while in local
The command is not executable while the device is in local due to a hard local control.
Example: The device receives a command which would change the rotary knob state,
but the device is in local so the command can not be executed.
-202
Settings lost due to rtl
A setting associated with hard local control was lost when the device changed to
LOCS from REMS or to LWLS from RWLS.
-210
Trigger error
Error on triggering the device.
-211
Trigger ignored
The trigger (GET, *TRG or trigger signal) was ignored because of device timing
considerations.
Example: The device was not ready to respond.
-212
Arm ignored
An arming signal was ignored by the device.
-213
Init ignored
Measurement initialization was ignored as another measurement was already in
progress.
Error code
Error text in the case of queue poll
Error explanation
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...