![Rohde & Schwarz R&S FSP Series Operating Manual Download Page 146](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864146.webp)
R&S FSP
Instrument Functions
Analyzer Mode
4.74
Operating Manual 1164.4556.12 - 06
Order of signals detected:
or
With
6 dB peak excursion
, all signals will be detected with
NEXT PEAK
and
NEXT
PEAK RIGHT
/
NEXT PEAK LEFT
.
Order of signals detected:
or
AUTO MAX
PEAK
AUTO MIN
PEAK
The
AUTO MAX PEAK
/
AUTO MIN PEAK
softkeys add an automatic peak search
action for marker 1 at the end of each particular sweep. This function may be used
during adjustments of a device under test to keep track of the current peak marker
position and level.
The current marker search limit settings (LEFT LIMIT, RIGHT LIMIT, THRESHOLD,
EXCLUDE LO) are taken into account.
Remote command:
CALC:MARK:MAX:AUTO ON | OFF
CALC:MARK:MIN:AUTO ON | OFF
PEAK:
Signal 1
NEXT PEAK:
Signal 2
NEXT PEAK:
Signal 2 (no further signal detected)
PEAK:
Signal 1
NEXT PEAK LEFT:
Signal 1 (no further signal detected)
NEXT PEAK RIGHT: Signal 2
NEXT PEAK RIGHT: Signal 2 (no further signal detected)
PEAK:
Signal 1
NEXT PEAK:
Signal 2
NEXT PEAK:
Signal 3
NEXT PEAK:
Signal 4
PEAK:
Signal 1
NEXT PEAK LEFT:
Signal 3
NEXT PEAK RIGHT: Signal 1
NEXT PEAK RIGHT: Signal 2
NEXT PEAK RIGHT. Signal 4
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...