External Generator Control – Option R&S FSP-B10
RECALL
The
RECALL
softkey restores the instrument setting with which the calibration was
carried out.
This can be useful if the device setting was changed after calibration (e.g. center fre-
quency setting, frequency deviation, reference level, etc.).
The softkey is only available if:
• the
NETWORK
mode has been selected
• the memory contains a calibration data set.
Remote command:
CORR:REC
SAVE AS TRD
FACTOR
SAVE AS TRD FACTOR uses the normalized measurement data to generate a
transducer factor with up to 625 points. The trace data are converted to a transducer
with unit dB after the transducer name has been entered. The number of points is
defined by SWEEP COUNT. The frequency points are allocated in equidistant steps
between start and stop frequency. The generated transducer factor can be further
adapted in the menu SETUP – TRANSDUCER. SAVE AS TRD FACTOR is only
available if normalization is switched on.
Remote command:
CORR:TRAN:GEN <name>'
4.9.3
Reflection Measurement
Scalar reflection measurements can be carried out by means of a reflection-coeffi-
cient bridge.
Fig. 4.34 Test setup for reflection measurement
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...