Operating Manual 1164.4556.12 - 06
4.153
R&S FSP
Instrument Functions
Basic Settings
Remote command:
CALC:LIM3:CONT:DOM FREQ
X-Axis
- Select interpolation
Linear or logarithmic interpolation can be carried out between the frequency refer-
ence points of the table. The ENTER key toggles between LIN and LOG selection.
Remote command:
CALC:LIM3:CONT:SPAC LIN
CALC:LIM3:UPP:SPAC LIN
CALC:LIM3:LOW:SPAC LIN
Scaling -
selection of absolute or relative scaling
The limit line can either be scaled in absolute (frequency or time) or relative units.
Any of the unit keys may be used to toggle between
ABSOLUTE
and
RELATIVE
,
the cursor must be positioned in the X
Scaling
or the
Y Scaling
line.
The
RELATIVE
scaling is always suitable, if masks for bursts are to be defined in the
time domain, or if masks for modulated signals are required in the frequency
domain.
An X offset with half the sweep time may be entered in order to shift the mask in the
time domain into the center of screen.
Remote command:
CALC:LIM3:CONT:MODE ABS
CALC:LIM3:UPP:MODE ABS
CALC:LIM3:LOW:MODE ABS
Unit
- Select the vertical scale units for the limit line
The selection of units takes place in a selection box. The default setting is
dBm
.
Remote command:
CALC:LIM3:UNIT DBM
Limit
- Select upper/lower limit
A limit line can be defined as either an upper or lower limit.
Remote command:
--
(defined by key words :UPPer or :LOWer)
A change in domain (frequency/time) is only permitted when the data point table is
empty.
X-Scaling
ABSOLUTE
The frequencies or times are interpreted as absolute
physical units.
X-Scaling
RELATIVE
In the data point table, the frequencies are referred to the
currently set center frequency. In time domain mode, the
left boundary of the diagram constitutes the reference.
Y-Scaling ABSOLUTE
The limit values refer to absolute levels or voltages.
Y-Scaling RELATIVE
The limit values refer to the reference level (Ref Level) or,
in case a reference line is set, to the reference line.
Limit values with the unit dB are always relative values.
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...