![Rohde & Schwarz R&S FSP Series Operating Manual Download Page 244](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864244.webp)
R&S FSP
Instrument Functions
Basic Settings
4.172
Operating Manual 1164.4556.12 - 06
Fig. 4.17 Linear frequency axis and linear interpolation
Fig. 4.18 Logarithmic frequency axis and interpolation
Comment -
Entry of comment
Any comment with a maximum length of 50 characters can be entered.
Remote command:
CORR:TRAN:COMM <string>
INSERT LINE
The
INSERT LINE
softkey inserts an empty line above the marked reference value.
When entering a new reference value in the line, the ascending order of frequencies
must be taken into consideration, however.
Remote command:
--
DELETE LINE
The
DELETE LINE
softkey deletes the marked reference value (complete line). The
reference values that follow move one line up.
Remote command:
--
SAVE TRD
FACTOR
The
SAVE TRD FACTOR
softkey saves the changed table in a file on the internal
hard disk.
If there is already a transducer factor that has the same name, a confirmation query
is output.
If the new factor is active, the new values become immediately valid.
Remote command:
--
(executed automatically after the definition of the
reference values)
4.6.3.5
Programming the Interface Configuration and Time Setup
The
GENERAL SETUP
softkey opens a submenu in which the general instrument
parameters can be set up. In addition to the configuration of the digital interfaces
(
IECBUS, COM
), the date and time may be entered.
The current settings are displayed in tabular form on the display screen where they
may be edited.
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...