![Rohde & Schwarz R&S FSP Series Operating Manual Download Page 290](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864290.webp)
R&S FSP
Instrument Functions
Tracking Generator – Option R&S FSP-B9
4.218
Operating Manual 1164.4556.12 - 06
Fig. 4.23 Measurement with REF VALUE -10 dB and REF VALUE POSITION 50%
After the reference line has been moved by entering a
REF VALUE
of –10 dB the
deviation from the nominal power level can be displayed with high resolution (e.g.
1 dB/div). The power is still displayed in absolute values, which means that in the
above example 1 dB below the nominal power (reference line) = 11 dB attenuation.
Fig. 4.24 Measurement of a 10dB attenuator pad with 1dB/DIV
Remote command:
DISP:WIND:TRAC:Y:RVAL -10dB