R&S FSP
Remote Control – Description of Commands
INPut Subsystem
6.126
Operating Manual 1164.4556.12 - 06
INPut<1|2>:GAIN:STATe
ON | OFF
This command switches the preamplifier for the instrument on or off. The switchable gain is fixed to 20
dB.
Example:
"INP:GAIN:STAT ON"
'Switches the 20 dB preamplifier on.
Characteristics:
*RST value: OFF
SCPI: conform
Mode:
A
The command is only available with the option electronic attenuator R&S FSP-B25.
INPut<1|2>:IMPedance
50 | 75
This command sets the nominal input impedance of the instrument. The set impedance is taken into
account in all level indications of results.
The setting 75
Ω
should be selected, if the 50
Ω
input impedance is transformed to a higher impedance
using a 75
Ω
adapter of the RAZ type (= 25
Ω
in series to the input impedance of the instrument). The
correction value in this case is 1.76 dB = 10 log (75
Ω
/ 50
Ω
).
Example:
"INP:IMP 75"
Characteristics:
*RST value: 50
Ω
SCPI: conform
Mode:
A
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...