![Rohde & Schwarz R&S FSP Series Operating Manual Download Page 778](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864778.webp)
R&S FSP
Error Messages
SCPI-Specific Error Messages
9.4
Operating Manual 1164.4556.12 - 06
-114
Header suffix out of range
The header contains an invalid numeric suffix.
Example:
SENSe3
does not exist in the instrument.
-120
Numeric data error
The command contains a faulty numeric parameter.
-121
Invalid character in number
A number contains an invalid character.
Example: An "A" in a decimal number or a "9" in an octal number.
-123
Exponent too large
The absolute value of the exponent is greater than 32000.
-124
Too many digits
The number includes too many digits.
-128
Numeric data not allowed
The command includes a number which is not allowed at this position.
Example: The command
INPut:COUPling
requires indication of a text parameter.
-130
Suffix error
The command contains a faulty suffix.
-131
Invalid suffix
The suffix is invalid for this instrument.
Example: nHz is not defined.
-134
Suffix too long
The suffix contains more than 12 characters.
-138
Suffix not allowed
A suffix is not allowed for this command or at this position of the command.
Example: The command
*RCL
does not permit a suffix to be indicated.
-140
Character data error
The command contains a faulty text parameter
-141
Invalid character data
The text parameter either contains an invalid character or it is invalid for this command.
Example: Write error with parameter indication;
INPut:COUPling XC
.
-144
Character data too long
The text parameter contains more than 12 characters.
-148
Character data not allowed
The text parameter is not allowed for this command or at this position of the command.
Example: The command
*RCL
requires a number to be indicated.
-150
String data error
The command contains a faulty string.
-151
Invalid string data
The command contains a faulty string.
Example: An END message has been received prior to the terminating apostrophe.
Error code
Error text in the case of queue poll
Error explanation
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...