![Rohde & Schwarz R&S FSP Series Operating Manual Download Page 780](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864780.webp)
R&S FSP
Error Messages
SCPI-Specific Error Messages
9.6
Operating Manual 1164.4556.12 - 06
-214
Trigger deadlock
The trigger source for the initiation of measurement is set to GET and subsequent
measurement is received. The measurement cannot be started until a GET is
received, but the GET would cause an interrupted-error)
-215
Arm deadlock
The trigger source for the initiation of measurement is set to GET and subsequent
measurement is received. The measurement cannot be started until a GET is
received, but the GET would cause an interrupted-error.
-220
Parameter error
The command contains a faulty or invalid parameter.
-221
Settings conflict
There is a conflict between setting of parameter value and instrument state.
-222
Data out of range
The parameter value lies out of the allowed range of the instrument.
-223
Too much data
The command contains too many data.
Example: The instrument does not have sufficient storage space.
-224
Illegal parameter value
The parameter value is invalid.
Example: The text parameter is invalid,
TRIGger:SWEep:SOURce TASTe
-230
Data corrupt or stale
The data are incomplete or invalid.
Example: The instrument has aborted a measurement.
-231
Data questionable
The measurement accuracy is suspect.
-240
Hardware error
The command cannot be executed due to problems with the instrument hardware.
-241
Hardware missing
Hardware is missing.
Example: An option is not fitted.
-250
Mass storage error
A mass storage error occurred.
-251
Missing mass storage
The mass storage is missing.
Example: An option is not installed.
-252
Missing media
The media is missing.
Example: There is no floppy in the floppy disk drive.
-253
Corrupt media
The media is corrupt.
Example: The floppy is bad or has the wrong format.
Error code
Error text in the case of queue poll
Error explanation
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...