![Rohde & Schwarz R&S FSP Series Operating Manual Download Page 469](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864469.webp)
Operating Manual 1164.4556.12 - 06
6.75
R&S FSP
Remote Control – Description of Commands
CALCulate Subsystem
CALCulate<1|2>:MARKer<1...4>:FUNCtion:POWer:SELect
ACPower | CPOWer | MCACpower |
OBANdwidth | OBWidth | CN | CN0
This command selects – and switches on – one of the above types of power measurement in the
selected measurement window. This function is independent of the selected marker, i.e. the numerical
suffix <1...4> appended to MARKer has no effect.
The channel spacings and channel bandwidths are configured in the
SENSe:POWer:ACHannel
subsystem.
Please note the following:
If CPOWer is selected, the number of adjacent channels (command:
[SENSe:]POWer:ACHannel:
ACPairs)
is set to 0. If ACPower is selected, the number of adjacent channels is set to 1, unless
adjacent channel power measurement is switched on already.
With respect to the above two settings, the behavior of the R&S FSP differs from that of the FSE family.
This command is a query and therefore has no *RST value.
The occupied bandwidth measurement is performed for the trace on which marker 1 is positioned. To
select another trace for the measurement, marker 1 is to be positioned on the desired trace by means
of
CALC:MARK:TRAC 1|2|3
.
Parameter:
ACPower: adjacent channel power measurement with a single carrier signal
CPOWer: Channel power measurement with a single carrier signal (equivalent to
adjacent channel power measurement with
NO. OF ADJ CHAN
= 0)
MCACpower: Channel/adjacent channel power measurement with several carrier
signals
OBANdwidth | OBWidth: Measurement of occupied bandwidth
CN: Measurement of carrier-to-noise ratio
CN0: Measurement of carrier-to-noise ratio referenced to 1 Hz bandwidth
Example:
"CALC:MARK:FUNC:POW:SEL ACP"
'Switches on adjacent channel power measurement in window A.
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
A-F
The channel/adjacent channel power measurement is performed for the trace
selected with
SENSe:POWer:TRACe 1|2|3
.
The parameters CN and CN0 are available only from firmware version 1.40.
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...