Command Reference
R&S
®
ZNB/ZNBT
950
User Manual 1173.9163.02 ─ 53
CONFigure:CHANnel<Ch>[:STATe]
<Boolean>
Creates or deletes channel no.
<Ch>
and selects it as the active channel.
defines the channel name.
A channel created using
CONFigure:CHANnel<Ch>[:STATe] ON
can be configured
but has no trace assigned so that no measurement can be initiated. Use
CALCulate<Ch>:PARameter:SDEFine
<TraceName>, <Result>
to create a
channel and a trace. In remote control, it is possible to remove all channels. This is in
contrast to manual control where at least one channel with one diagram area and one
trace must be available.
Suffix:
<Ch>
.
Number of the channel to be created or deleted.
Parameters:
<Boolean>
ON - create channel no.
<Ch>
. If the channel no.
<Ch>
exists
already, it is not modified but selected as the active channel.
OFF - delete channel no.
<Ch>
.
*RST:
ON for channel no. 1 (created on *RST), OFF for all
other channels.
Example:
See
Manual operation:
See
CONFigure:CHANnel<Ch>:MEASure[:STATe]
<Boolean>
Enables or disables the sweep in channel no.
<Ch>
. This command can be used to
restrict the measurement in a subset of channels in order to optimize the measurement
speed.
Suffix:
<Ch>
.
Number of an existing channel.
Parameters:
<Boolean>
ON | OFF
*RST:
ON (all existing channels)
Example:
*RST; :CONFigure:CHANnel2 ON; CHANnel3 ON
Create channels 2 and 3, in addition to the default channel no.
1. The analyzer performs sweeps in all three channels.
CONFigure:CHANnel:MEASure:ALL OFF
Disable the measurement in all channels
CONFigure:CHANnel2:MEASure ON
(Re-)enable the measurement in channel no. 2. The analyzer
measures in channel 2; the channels no. 1 and 3 are not mea-
sured.
Manual operation:
See
SCPI Command Reference