![Rohde & Schwarz 1311.6010.22 User Manual Download Page 506](http://html.mh-extra.com/html/rohde-and-schwarz/1311-6010-22/1311-6010-22_user-manual_1477768506.webp)
GUI Reference
R&S
®
ZNB/ZNBT
506
User Manual 1173.9163.02 ─ 53
Remote command:
OFF
Channel On
Toggles the measurement state of the
.
Remote command:
CONFigure:CHANnel<Ch>:MEASure[:STATe]
Fixture Simulator
The "Fixture Simulator" switch deactivates or activates the configured deembedding,
embedding, balanced ports, and port impedance settings for the selected channel.
When "Fixture Simulator" is
deactivated:
●
all balanced ports are resolved to single ended ports
●
all port impedances are set to default
●
all de/embeddings are disabled
At the GUI, the "Balanced Ports" dialog and the de/embedding tabs and dock widgets
are disabled. Related remote commands generate an error.
When "Fixture Simulator" is
reactivated
, the situation before the deactivation is
restored.
Note:
The "Offset" and "One Way Loss" settings are
not
affected by the "Fixture Simu-
lator" switch. Use the "All Compensation Off"/"All Compensation On" functions of the
"Fixture Compensation" dialog to de/activate these compensations as well (see
ter 6.13.2.2, "Fixture Compensation Dialog"
Remote command:
CALCulate<Ch>:TRANsform:VNETworks:FSIMulator[:STATe]
6.12.1.2
Channel Manager Dialog
The "Channel Manager" dialog allows you to rename channels and to change their
measurement state and sweep mode.
Access:
CHANNEL – [CHANNEL
CONFIG] > "Channels" > "Channel Manager..."
Background information
Refer to
Chapter 5.1.3.3, "Active and Inactive Traces and Channels"
Channel Config Softtool