Command Reference
R&S
®
ZNB/ZNBT
763
User Manual 1173.9163.02 ─ 53
8.3.1.4
CALCulate:DTIMe...
Defines the properties and retrieves the results of the skew measurement provided
with the
option R&S
ZNB/ZNBT-K20.
..................................................................................... 763
CALCulate<Chn>:DTIMe:LIMit:FAIL?
............................................................................... 763
CALCulate<Chn>:DTIMe:LIMit:FAIL:BEEP
....................................................................... 764
CALCulate<Chn>:DTIMe:LIMit:LIMit
................................................................................ 764
CALCulate<Chn>:DTIMe:LIMit:STATe
.............................................................................. 764
...................................................................................765
...................................................................................... 765
....................................................................................765
CALCulate<Chn>:DTIMe:DATA?
[<Data>]
Queries the results of the skew measurement
Suffix:
<Chn>
.
Channel number used to identify the active trace
Query parameters:
<Data>
ALL
If omitted, a single numeric value is returned. If ALL is specified,
the result consists of 6 numeric values. Furthermore, the inter-
pretation of the result values depends on the active trace's stim-
ulus axis (
CALCulate<Chn>:TRANsform:TIME:XAXis
). See
the table below.
Note
that the skew value is positive, if, at the defined position
(
the right of the reference trace.
Usage:
Query only
Options:
R&S
ZNB/ZNBT-K20
Manual operation:
See
<Data> omitted
<Data> = ALL
Stimulus axis:
time
<skew time>
<skew time>,<skew distance>, <X value of ref. point on current
trace>, <Y value of ref. point on current trace>, <X value of ref.
point on reference trace>, <Y value of ref. point on reference
trace>
Stimulus axis:
distance
<skew distance>
<skew distance>, <skew time>, <X value of ref. point on cur-
rent trace>, <Y value of ref. point on current trace>, <X value of
ref. point on reference trace>, <Y value of ref. point on refer-
ence trace>
CALCulate<Chn>:DTIMe:LIMit:FAIL?
<Boolean>
Indicates whether the skew has passed or failed.
SCPI Command Reference