GUI Reference
R&S
®
ZNB/ZNBT
477
User Manual 1173.9163.02 ─ 53
the balanced port was dissolved and only the uncalibrated (single-ended) port was
disabled.
In earlier FW versions there was no such port deactivation mechanism.
●
Since FW version 2.40 an uncalibrated port is only disabled if it is not used by a
measurement, i.e. if it is not required by any trace of the related channel.
Remote command:
The port parameters in many calibration commands define ports to be calibrated.
Type
Selects the calibration type to be performed. For background information, see
ter 5.5.7.2, "Calibration Types"
SMARTerCal is based on a full n-port calibration (n > 1): "Adapter Removal" is a 2-port
calibration, all other calibration types can be used for any combination of 2 or more
ports.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
Power Meter
Shows a list of all power meters that have been properly configured and are available
for the scalar source power calibration of a single source port. The last configured
power meter is selected by default. See
Remote command:
[SENSe:]CORRection:COLLect:PMETer:ID
Step 2: Connectors and Cal Kits
Selects the connector type and gender for all ports and allows you to import a calibra-
tion kit.
The upper part of the panel shows the previously configured SMARTerCal(s). Select
the appropriate one to get access to the related port settings. For the individual calibra-
tions, the functionality is the same as described for the single calibration version of the
wizard (see
"Step 2: Connectors and Cal Kits"
Cal Softtool