GUI Reference
R&S
®
ZNB/ZNBT
353
User Manual 1173.9163.02 ─ 53
6.6.2.1
Controls on the Ripple Test Tab
The "Def. Ripple Test..." button opens the "Define Ripple Test" dialog (see
ter 6.6.2.2, "Define Ripple Test Dialog"
Show Ripple Limits
Shows or hides the ripple limit lines associated with the active trace in a Cartesian dia-
gram area. The vertical positions of the ripple lines are recalculated after each sweep;
only their stimulus range and distance (the ripple limit) are fixed.
Note:
Display of the limit line and limit check are independent of each other: Hiding the
limit line does not switch off the limit check.
Remote command:
CALCulate<Chn>:RIPPle:DISPlay[:STATe]
Ripple Check
Switches the ripple limit check of the active trace on or off.
When the limit check is switched on, a movable info field shows the pass/fail informa-
tion and the measured ripple in each ripple limit range. If the ripple limit check fails at a
measurement point, the point is marked with a colored square.
The "Limit Fail Trace Color" is defined in the
Define User Color Scheme Dialog
.
Lines Softtool