GUI Reference
R&S
®
ZNB/ZNBT
675
User Manual 1173.9163.02 ─ 53
Auto Averaging
Activates automatic averaging, which means that the VNA performs multiple calibration
sweeps and applies averaging to reduce trace noise. In contrast to regular averaging
(see
on page 391), the number of calibration sweeps is
calculated automatically.
Remote command:
[SENSe:]CORRection:COLLect:AVERage
Multiple Cal in Calibration Wizard
Enables/disables multiple calibrations in the calibration wizard, which allows to set up
and perform multiple calibrations in one manual calibration procedure. See
ter 6.11.1.4, "Calibration Unit Wizard (MultiCal)"
bration Setting Wizard (MultiCal)"
Chapter 6.11.1.9, "SMARTerCal Wiz-
Chapter 6.11.1.11, "SMARTerCal Wizard
Remote command:
SYSTem:DISPlay:DIALogs:SETup:MCAL[:STATe]
Show Cal Kit Label
Enables/disables the "Calibration Info" dialog during manual calibration (see
Independent of the state of the "Show Cal Kit Label" flag, cal kit labels are displayed in
several other manual calibration dialogs.
Use Reduced Number of Through
logic for every port assignment of an auto-
matic calibration.
Note that for this to work, the port assignments must have a star topology, with a
"Common Port" set using
[SENSe<Ch>:]CORRection:COLLect:AUTO:CPORt
Remote command:
[SENSe:]CORRection:COLLect:AUTO:MCONnect
Delete Cal Pool / Delete All Cal Kits
Deletes all calibration data and all cal kit data. See
Chapter 6.11.4.3, "Calibration Man-
Remote command:
n/a
Search Path for additional Cal Kits and Connector Types
Contains the name and path of a special directory for cal kit files (
*.calkit
). All cal
kit files in the special directory are loaded automatically as predefined kits (i.e. read-
only kits which cannot be modified) every time the VNA application is started. It is pos-
sible to select the default cal kit directory
C:\Users\Public\Documents\Rohde-Schwarz\Vna\Calibration\Kits
or
any other directory. "None" means that no additional cal kit files are loaded on start-up.
Setup Softtool