GUI Reference
R&S
®
ZNB/ZNBT
699
User Manual 1173.9163.02 ─ 53
●
"Interface" selects an interface/protocol type for the connection. In addition to the
GPIB, VXI-11, SOCKET, and USB-VISA interface types (for devices connected to
the GPIB Bus, LAN or USB connectors of the analyzer; see
), the ana-
lyzer supports any "Other" interface supported by the installed VISA library.
●
"Address" contains the address for the current interface type. GPIB addresses
must be unique for all devices connected to the GPIB bus (range: 0 to 30), GPIB
and IP addresses must agree with the entries in the VISA library. The remaining
interface types require composite address formats; see
If an instrument is connected to the R&S
ZNB/ZNBT, the entries in the [Driver for
New Instrument] panel can be auto-detected for the specified interface type and
address.
●
"Identify" sends an identification query ("IDN?") to the specified device address in
order to identify the type and serial number of the connected generator and select
an appropriate driver file. Generator driver files (
*.gen
) are stored in the
Resources\ExtDev
subdirectory of the analyzer's program directory.
Table 6-8: Interface types for external generators and address formats
Physical
interface
(connector)
Interface
(protocol)
Address
Remarks
LAN
VXI-11
<IpAddress>
e.g.
127.0.0.0
Full VISA resource string:
TCPIP[board]::<Address>[::INSTR]
SOCKET
<IpAddress>::<PortNo>
e.g.
127.0.0.0::50000
LAN connection with pure TCP/IP protocol;
refer to your VISA user documentation.
GPIB
GPIB0 ...
GPIB9
<Address>
e.g.
20
Full VISA resource string:
GPIB[board]::<Address>[::INSTR]
Setup Softtool