Command Reference
R&S
®
ZNB/ZNBT
1187
User Manual 1173.9163.02 ─ 53
Parameters:
<IntSourcePortPow> Power level for port <PortId> in sweep segment <Seg> of chan-
nel <Ch>
Default unit: dBm
Manual operation:
See
[SENSe<Ch>:]SEGMent<Seg>:PORT<PortId>[:STATe]
<Boolean>
Enables a port-specific power level (base power) for port <PortId> in segment <Seg>
of channel <Ch>.
Suffix:
<Ch>
.
Channel number
<Seg>
Sweep segment number
<PortId>
Port number
Parameters:
<Boolean>
OFF (0)
No port-specific power level for port <PortId> in segment <Seg>
of channel <Ch>.
If
[SENSe<Ch>:]SEGMent<Seg>:POWer[:LEVel]:CONTrol
is set to ON, the segment power level
LEVel][:IMMediate][:AMPLitude]
ON (1)
The port-specific power level
segment <Seg> of channel <Ch>.
*RST:
OFF
Manual operation:
See
[SENSe<Ch>:]SEGMent<Seg>:POWer:GAINcontrol
<ReceiverName>, <Mode>
[SENSe<Ch>:]SEGMent<Seg>:POWer:GAINcontrol?
<ReceiverName>
Defines port- and segment-specific IF gain settings.
These settings apply if and only if segmented gain control (GC) is enabled (see
[SENSe<Ch>:]SEGMent<Seg>:POWer:GAINcontrol:CONTrol
).
Note:
●
Without the
option only the gain of the measurement
receivers (the b-waves) can be statically set (to LNOise or LDIStortion); the refer-
ence receiver always uses automatic gain control (AUTO). With the option availa-
ble, also the gain of the reference receivers (the a-waves) can be set statically.
●
instead, if
no distinction between different sweep segments shall be made.
SCPI Command Reference