2. Perform a SMARTerCal:
Figure 6-45: System Error Correction
Figure 6-46: Scalar Power Calibration
6.13
Offset Embed Softtool
The "Offset Embed" softtool allows you to define a length offset and loss for each test
port. The offset compensates for the known length and loss of (non-dispersive and per-
fectly matched) transmission lines between the calibrated reference plane and the