Command Reference
R&S
®
ZNB/ZNBT
1142
User Manual 1173.9163.02 ─ 53
Manual operation:
See
Test Port
Port Assignment 1
Port Assignment 2
2
Cal Unit Port 1
-
3
Cal Unit Port 2
auto-detected
4
-
auto-detected
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:DEFine:TPORt
<TestPort1>, <TestPort2>...
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:
, but initially defines the port assignment only by its underlying test port set; the
connected calibration unit ports are
auto-detected
at the start of the corresponding
calibration sweep.
Suffix:
<Ch>
.
Number of the channel
<Asg>
Number of the port assignment
Parameters:
<TestPort1>
First test port number
<TestPort2>
Second test port number ...
Example:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:SAVE
Calculates the system error correction for automatic calibrations with
, saves the data and applies the calibration to the active channel.
Requires successful calibration sweeps for all related port assignments (see
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:ACQuire
)
Suffix:
<Ch>
.
Channel number of the calibrated channel
Example:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:
Usage:
Event
Manual operation:
See
[SENSe<Chn>:]CORRection:PSTate?
Gets the power calibration label of the active trace in channel <Chn>.
SCPI Command Reference