Command Reference
R&S
®
ZNB/ZNBT
895
User Manual 1173.9163.02 ─ 53
Suffix:
<Chn>
.
Channel number
This suffix is ignored: cable types are defined for (and deleted
from) all channels.
Setting parameters:
<DtfDeleteCable>
Name of a user-defined cable type.
Example:
CALCulate:TRANsform:DTFault:DELete 'My cable
type'
Deletes the user-defined cable type "My cable type".
Usage:
Setting only
Options:
R&S
ZNB/ZNBT-K3
Manual operation:
See
CALCulate<Chn>:TRANsform:DTFault:SELect
<DtfSelectCable>
Selects one of the available (predefined or user-defined) cable types for Distance to
Fault measurements.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Parameters:
<DtfSelectCable>
Name of the cable type.
Example:
CALCulate1:TRANsform:DTFault:SELect '5088-HFLR'
Selects cable type "5088-HFLR".
Manual operation:
See
CALCulate<Chn>:TRANsform:DTFault:CENTer
<Center>
CALCulate<Chn>:TRANsform:DTFault:SPAN
<Span>
Defines the distance window of the Distance to Fault measurement using its center
and span.
See also
CALCulate<Chn>:TRANsform:DTFault:STARt
.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Parameters:
<Center>/<Span>
Center/span value.
Default unit: m
Example:
:CALCulate1:TRANsform:DTFault:CENTer 2; SPAN 2
Defines the DtF distance window of the active trace of channel 1
via its center (2m) and span (2m).
Options:
R&S
ZNB/ZNBT-K3
SCPI Command Reference