Coupon Type
Selects the coupon type to be measured. The following coupon types are supported:
●
"Sym 2x Thru"
●
"1x Open"
●
"1x Short"
●
"1x Open, 1x Short" (ISD only)
Remote command:
CALCulate:FMODel:ISD<Ph_pt>:COUPon:TYPE
CALCulate:FMODel:SFD<Ph_pt>:COUPon:TYPE
Measure / Measure Open / Measure Short
Starts the coupon measurement at the
ports.
The display will change to show all the S parameter measurements being made. The
resulting Touchstone file is written to C:\Users\Public\Documents\Rohde-Schwarz\Vna
\Embedding.
For
"1x Open, 1x Short" (ISD only) the measurement proceeds in two
steps: one for the "1x Open" and one for the "1x Short" coupon. The label of the "Mea-
sure" button changes accordingly.
Remote command:
CALCulate:FMODel:ISD<Ph_pt>:COUPon:MEASure
CALCulate:FMODel:ISD<Ph_pt>:COUPon:MEASure:OPEN
CALCulate:FMODel:ISD<Ph_pt>:COUPon:MEASure:SHORt
CALCulate:FMODel:SFD<Ph_pt>:COUPon:MEASure
Load File / 1x Open Preset / 1x Short Preset
Allows you to load the coupon properties from a Touchstone file.
Table 6-5: Allowed Touchstone file types
"Sym 2x Thru"
other
Single Ended
s2p
s1p
Balanced
s2p or s4p
s2p
Offset Embed Softtool