![Rohde & Schwarz 1311.6010.22 User Manual Download Page 765](http://html.mh-extra.com/html/rohde-and-schwarz/1311-6010-22/1311-6010-22_user-manual_1477768765.webp)
Command Reference
R&S
®
ZNB/ZNBT
765
User Manual 1173.9163.02 ─ 53
Manual operation:
See
CALCulate<Chn>:DTIMe:POSition
<SkewPos>
Defines the position of the skew measurement.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<SkewPos>
Skew position as integer percentage of the step size
Options:
R&S
ZNB/ZNBT-K20
Manual operation:
See
CALCulate<Chn>:DTIMe:STATe
<Boolean>
Enables/disables the
.
Note:
The skew measurement can only be performed if the following conditions are
met for both the active trace and the reference trace:
●
REAL
)
●
Time Domain is enabled (
CALCulate<Chn>:TRANsform:TIME:STATe
ON
)
●
the Low Pass Step time domain transform is used
LPASs
CALCulate<Chn>:TRANsform:TIME:
STEP
)
The latter, in turn, requires the stimulus grid to be harmonic. This can be achieved, for
example, using
ON
.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean>
*RST:
OFF
Options:
R&S
ZNB/ZNBT-K20
Manual operation:
See
CALCulate<Chn>:DTIMe:TARGet
<SkewRefTrace>
Selects the reference trace for the skew measurement.
See
for conditions on both the active and the refer-
ence trace.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<SkewRefTrace>
Reference trace, identified by its name (within channel <Chn>).
SCPI Command Reference