Channel table
The rows and columns of the channel table represent the existing channels (rows)
together with certain editable (white) or non-editable (gray) properties (columns).
●
"Name" indicates the name of the related channel.
●
"Traces" indicates the names of all traces assigned to the related channel.
●
"On/Off" toggles the measurement state of the related channel.
●
"Single Sweep" toggles between "Continuous" and "Single" sweep mode (see
Remote command:
CONFigure:CHANnel:CATalog?
CONFigure:CHANnel<Ch>:NAME
CONFigure:CHANnel<Ch>:NAME:ID?
CONFigure:CHANnel<Ch>:MEASure[:STATe]
INITiate<Ch>:CONTinuous
Add / Delete
The buttons below the channel table add and delete channels.
●
"Add" adds a new channel to the list. The new channel is named Ch<n>, where
<n> is the largest of all existing channel numbers plus one.
●
"Delete" deletes the channel selected in the table. This button is disabled if the
setup contains only one channel: In manual control, each setup must contain at
least one diagram area with one channel and one trace.
Remote command:
6.12.1.3
RFFE Config Dialog for R&S
ZN-B15/-Z15 Var. 03
Controls the channel-specific setup of the two RFFE bus interfaces and GPIO ports
provided by
variant 03
of the internal/external RFFE GPIO extension board R&S
ZN-
B15/-Z15.
Access:
CHANNEL – [CHANNEL
CONFIG] > "Channels" > "RFFE..."
Channel Config Softtool