GUI Reference
R&S
®
ZNB/ZNBT
410
User Manual 1173.9163.02 ─ 53
●
"Point" means that each trigger event starts the measurement at the next sweep
point.
●
"Partial Measurement" means that each trigger event starts the next partial mea-
surement at the current or at the next sweep point. If every sweep point only
requires a single partial measurement, this option is equivalent to "Point".
See also
Chapter 5.1.4.1, "Partial Measurements and Driving Mode"
●
"Segment" means that each trigger event starts the next sweep segment within the
current sweep. If a sweep type other than
is active, this option is equiv-
alent to "Sweep".
Relation with other sweep settings
Some sweep settings are logically incompatible with a particular selection of the trig-
gered measurement sequence:
●
If a
sweep is performed, the sequence is always a sweep.
●
"Alternated"
only makes sense if the triggered measurement
sequence comprises more than one sweep point. If "Point" or "Partial Measure-
ment" is selected, "Alternated" mode is switched off and vice versa.
Note:
The trigger events must be adjusted to the triggered measurement sequence. If
the analyzer receives a trigger event while the last sequence is still running, the
R&S
ZNB/ZNBT skips the trigger event and generates a message.
Remote command:
Delay
Specifies a delay time between the trigger event and the start of the next measurement
sequence.
The specified "Delay" must be zero or positive, so that the trigger event precedes the
start of the measurement (post-trigger).
If "Multiple Triggers" is active, the "Delay" can be selected individually using the
.
Remote command:
TRIGger<Ch>[:SEQuence]:HOLDoff
Signal Type
Specifies the "External" trigger mode in detail.
●
"Rising Edge"/"Falling Edge"
means that the rising/falling slope of every external
trigger pulse can trigger a single measurement sequence.
●
"Active High"/"Active Low"
means that the analyzer measures in "FreeRun"
mode as long as the external trigger signal is high/low. The measurement is dis-
continued when the trigger signal changes to low/high.
Remote command:
6.10.3.2
Trigger Manager Dialog
The "Trigger Manager" dialog defines individual trigger sources and delays for the trig-
gered measurement sequences. The dialog is available and its settings are valid if the
Sweep Softtool