Command Reference
R&S
®
ZNB/ZNBT
1034
User Manual 1173.9163.02 ─ 53
Manual operation:
See
"Add / Add All... / Replace / Apply / Apply to All"
MMEMory:LOAD:CORRection:RESolve
<Channel>[, <CalGroupFile>]
Resolves the pool link between channel
<Channel>
and a correction data set (cal
group file). After resolving the pool link, the analyzer keeps the previous system error
correction as a channel calibration ("Channel Cal"). A new calibration will replace the
channel calibration but not overwrite the old cal group file (and not affect other chan-
nels).
Setting parameters:
<Channel>
Channel number of an existing channel.
ALL
resolves the pool
link for all channels.
<CalGroupFile>
Optional string parameter to specify the name of the cal group
file. Cal group files must have the extension
*.cal
. The direc-
tory must not be specified; the analyzer always uses the default
path
C:\Users\Public\Documents\Rohde-Schwarz\Vna
\Calibration\Data
. If there is no link between
<Ch>
and the
specified file, the command has no effect. If no file is specified,
the command resolves any link between
<Ch>
and an arbitrary
cal group file.
Example:
See
Usage:
Setting only
Manual operation:
See
"Resolve Pool Link / Remove Pool Link"
MMEMory:LOAD:CORRection:TCOefficient<Ch>
<TraceFile>[, <Trace>]
Loads power meter (two-port) transmission coefficients from the specified power meter
correction file or trace file to channel <Ch>.
Suffix:
<Ch>
.
Channel number
Setting parameters:
<TraceFile>
String parameter specifying the name and directory of the loa-
ded trace file. The R&S
ZNB/ZNBT supports power meter cor-
rection list files (
*.pmcl
, generated using
*csv
, and Touchstone
(
*.s1p, *.s2p, ...
) files.
If no path is specified the analyzer searches the current direc-
tory, to be queried with
The file extensions
*.s<n>p
,
*csv
, and
*.pmcl
for Touch-
stone, ASCII, and power meter correction list files are manda-
tory.
*RST:
n/a
SCPI Command Reference