GUI Reference
R&S
®
ZNB/ZNBT
707
User Manual 1173.9163.02 ─ 53
The definition of the RF connections typically involves the following steps:
1. "Add" one or more switching matrices
These matrices must have been registered before; see
on page 701 for details.
2. For every matrix VNA port (the "north" ports in the graphical representation), spec-
ify the VNA port to which this matrix VNA port is physically connected or set it to
"Unused" in case no such connection exists.
3. For some matrix test ports and every VNA port that is not connected to a matrix,
assign a test port
Add
Opens the "Add Switch Matrix" dialog:
Select the required matrices and tap "OK" to add them to the RF connection configura-
tion or tap "Cancel" to quit the dialog without adding a switch matrix to the RF connec-
tion configuration
Note:
●
This dialog is shown automatically, in case no matrix has been added to the RF
connection configuration yet.
●
If the RF configuration already contains all registered switch matrices, the "Add"
button is disabled.
Remote command:
n.a.
Delete
Opens the "Delete Switch Matrix" dialog:
Setup Softtool