GUI Reference
R&S
®
ZNB/ZNBT
413
User Manual 1173.9163.02 ─ 53
Continuous / Single
Activate either continuous or single sweep mode for the active channel.
●
In "Continuous" mode, the analyzer measures continuously, repeating the current
sweep.
●
In "Single" sweep mode, the measurement is stopped after the configured number
of
initiates a new measurement cycle.
Tip:
Use "All Channels Continuous" or "All Channels on Hold" to select a common sweep
mode for all channels.
other than DEFAULT is used and "Sweep All Channels" is
selected in the
, then the selected sweep mode applies to all
channels.
Remote command:
See also:
CONFigure:CHANnel<Ch>:MEASure[:STATe]
CONFigure:CHANnel:MEASure:ALL[:STATe]
Sweeps
Selects the number of sweeps to be measured in "Single" sweep mode: the measure-
ment is stopped after the specified number of sweeps. This setting applies to the active
channel.
Tip:
other than DEFAULT is used and "Sweep All Channels" is
selected in the
, then "Sweeps" applies to all channels.
Remote command:
[SENSe<Ch>:]SWEep:COUNt
[SENSe:]SWEep:COUNt:ALL
Restart Sweep
Stops the current measurement and restarts a measurement cycle. In "Single" sweep
mode, a new single sweep sequence is started.
Remote command:
INITiate<Ch>[:IMMediate][:DUMMy]
INITiate[:IMMediate]:ALL
Restart Manager...
other than DEFAULT is selected (i.e. if some compatibility mode
is active), this button opens the
. Otherwise it is inactive.
All Channels Continuous / All Channels on Hold
Selects a common sweep control mode for all channels of the active recall set.
●
"All Channels Continuous": The R&S
ZNB/ZNBT continuously repeats the sweeps
in all channels.
●
"All Channels on Hold": The R&S
ZNB/ZNBT performs single sweeps, according to
the channel-specific number of "Sweeps".
Sweep Softtool